Sensitization to methyl methacrylate in the plastic catheter of an insulin pump infusion set

Saccabusi, Stanislao and Boatto, Gianpiero and Asproni, Battistina and Pau, Amedeo (2001) Sensitization to methyl methacrylate in the plastic catheter of an insulin pump infusion set. Contact Dermatitis, Vol. 45 (1), p. 47-48. ISSN 0105-1873. Article.

Full text not available from this repository.

Alternative URLs:

Abstract

Cases of allergic contact dermatitis due to acrylics in the adhesive used to attach the needle to the tubing have previously been reported in pump infusion set users, but never previously due to an acrylic in the plastic tubing itself, as in this case.
The presence of methyl methacrylate demonstrated in the catheter of the patient’s own infusion set correlated with her positive patch test results. Since patch testing and chemical-physical analysis was negative for methyl methacrylate in the alternative Cliniset® microinfusion set, this was supplied to the patient for her insulin therapy and no further adverse skin reactions were reported.
Patch testing demonstrated probable cross-sensitivity between methyl methacrylate and other acrylics.

Item Type:Article
ID Code:5222
Status:Published
Refereed:Yes
Uncontrolled Keywords:(meth)acrylates, allergic contact dermatitis, methyl methacrylate, insulin pump infusion, cross-reactivity, capillary gas-chromatography with mass quadrupole detector
Subjects:Area 03 - Scienze chimiche > CHIM/08 Chimica farmaceutica
Divisions:001 Università di Sassari > 01 Dipartimenti > Farmaco, chimico, tossicologico
Publisher:Munksgaard / Wiley
ISSN:0105-1873
Copyright Holders:© Munksgaard 2001
Deposited On:10 Dec 2010 16:28

I documenti depositati in UnissResearch sono protetti dalle leggi che regolano il diritto d'autore

Repository Staff Only: item control page