Saccabusi, Stanislao and Boatto, Gianpiero and Asproni, Battistina and Pau, Amedeo (2001) Sensitization to methyl methacrylate in the plastic catheter of an insulin pump infusion set. Contact Dermatitis, Vol. 45 (1), p. 47-48. ISSN 0105-1873. Article. Full text not available from this repository. Alternative URLs: AbstractCases of allergic contact dermatitis due to acrylics in the
adhesive used to attach the needle to the tubing have
previously been reported in pump infusion set users, but never previously due to an acrylic in the plastic
tubing itself, as in this case.